Silicone oil standards are recommended for the routine calibration of Viscometers made by:
Note: For Brookfield the last letters F or T indicates 4 or 8 speeds. Viscometers UK are self-explanatory.
SPRING TYPE | RECOMMENDED STANDARD | TEMPERATURE | SPINDLE | SPEED RPM |
---|---|---|---|---|
LV Series | 1000 cPs Cat No. RT1000 | 25.0 C | L3 | 60 |
RV Series | 5000 cPs Cat No. RT5000 | 25.0 C | R5 | 50 |
HV Series | 12500 cPs Cat No.RT12500 | 25.0 C | R3 | 50 |
Silicone oil standards are recommended for the routine calibration of Viscometers made by:
Note: For Brookfield the last letters F or T indicates 4 or 8 speeds. Viscometers UK are self-explanatory.
Spring Type | Recommended Standard | Temperature | Spindle | Speed rpm |
---|---|---|---|---|
LV Series | 1000 cPs Cat No. RT1000 | 25.0 C | L3 | 60 |
RV Series | 5000 cPs Cat No. RT5000 | 25.0 C | R5 | 50 |
HV Series | 12500 cPs Cat No.RT12500 | 25.0 C | R3 | 50 |
NOTE: With disc type spindles it is sometimes necessary to tilt the spindle slightly while immersing to avoid trapping air bubbles on the disc under surface. You may find it more convenient to immerse the spindle in this fashion before attaching it to the viscometer.
If necessary, calibration can also be checked at other points. Listings of further calibration sets can be given.
Your instruments reading should be within 1% +/- of the full scale reading. Affix new calibration label and date label. Labels are available from QC Labs.
Please also remember that the standard is calibrated to 0.2% of its stated value and you will need to allow for this possible error as well.
NOTE: If your reading is outside these error tolerances then you should check the following:
If your reading is still outside the error limits then it may need servicing and re-calibration. To arrange please contact:
Quality Control Laboratory13A Newbury RdChingford
London
E4 9JH
Fax: 020 8523 3003
If you have any problems carrying out your calibration check, or require information on the calibration of any other types of viscometers, please contact us at the above address.
The Calibration oils used in the above procedure are accurate to +/- 0.2% and are certified over 6 temperatures from 20.00 to 27.00 Degrees Cenntigrade.
A listing of all other calibration standards can be seen here.